Items 0
5350B
 
 

5350B

 

Profibus PA / Foundation Fieldbus transmitter

 

  • PROFIBUS PA ver. 3.0
  • FOUNDATION Fieldbus ver. ITK 4.6
  • Automatic switch between protocols
  • FISCO-certified
  • Basic or LAS capability with F.F.
  1. Overview +

  2. Technical details +

  3. Accessories +


Application

  • Linearized temperature measurement with RTD or TC sensor.
  • Difference, average or redundancy temperature measurement with RTD or TC sensor.
  • Linear resistance, potentiometer and bipolar mV measurement.

Technical characteristics

  • Bus transmitter with both PROFIBUS PA and FOUNDATION Fieldbus communication. A unique switch function ensures automatic shift between the two protocols.
  • Set-up for PROFIBUS PA can be done via Siemens Simatic® PDM®, ABB Melody / Harmony and Metso DNA software and for FOUNDATION Fieldbus via Emerson DeltaV, Yokogawa CS 1000 / CS 3000, ABB Melody / Harmony and Honeywell Experion software.
  • The simulation mode function can be activated by way of a magnet.
  • Polarity-independent bus connection.
  • 24 bit A/D converter ensures high resolution.
  • PROFIBUS PA function blocks:
    2 analog.
  • FOUNDATION Fieldbus function blocks:
    2 analog and 1 PID.
  • FOUNDATION Fieldbus capability:
    Basic or LAS.

Mounting / installation

  • For DIN form B sensor head mounting.

Environmental Conditions

  • Operating temperature

    -40°C to +85°C

  • Calibration temperature

    20...28°C

  • Relative humidity

    < 95% RH (non-cond.)

  • Protection degree (encl./terminal)

    IP68 / IP00

Mechanical specifications

  • Dimensions

    Ø 44 x 20.2 mm

  • Weight approx.

    55 g

  • Screw terminal torque

    0.4 Nm

  • Vibration

    DIN class B, IEC 60068-2-6 and IEC 60068-2-64

  • 2...25 Hz

    ±1.6 mm

  • 25...100 Hz

    ±4 g

Common specifications

  • Supply

  • Supply voltage

    9.0...30 VDC

  • Supply voltage in FISCO installations

    9.0...17.5 VDC

  • Max. required power

    < 350 mW

  • Isolation voltage

  • Test voltage

    1.5 kVAC for 60 s

  • Working voltage

    50 VRMS / 75 VDC

  • Response time

  • Response time (programmable)

    1...60 s

  • Quiescent current

    < 11 mA

  • Max. current increase in the event of an error

    < 7 mA

  • Warm-up time

    30 s

  • Signal / noise ratio

    Min. 60 dB

  • Updating time

    < 400 ms

  • Execution time, analog input

    < 50 ms

  • Accuracy

    Better than 0.05% of selected range

  • Signal dynamics, input

    24 bit

  • EMC immunity influence

    < ±0.1% of reading

  • Extended EMC immunity: NAMUR NE21, A criterion, burst

    < ±1% of reading

Input specifications

  • RTD input

  • RTD type

    Pt25…1000, Ni25…1000, Cu10…1000, lin. R, potentiometer

  • Cable resistance per wire

    50 Ω (max.)

  • Sensor current

    Nom. 0.2 mA

  • Effect of sensor cable resistance (3-/4-wire)

    < 0.002 Ω / Ω

  • Sensor error detection

    Yes

  • Short circuit detection

    < 15 Ω

  • TC input

  • Thermocouple type

    B, E, J, K, L, N, R, S, T, U, W3, W5

  • Cold junction compensation (CJC)

    < ±0.5°C

  • Sensor error detection

    Yes

  • Sensor error current: When detecting / else

    Nom. 4 μA / 0 μA

  • Short circuit detection

    < 3 mV

  • Voltage input

  • Measurement range

    -800...+800 mV

  • Input resistance

    10 MΩ

Output specifications

  • PROFIBUS PA connection

  • PROFIBUS PA protocol

    Profile A&B, ver. 3.0

  • PROFIBUS PA protocol standard

    EN 50170 vol. 2

  • PROFIBUS PA address (at delivery)

    126

  • PROFIBUS PA function blocks

    2 analog

  • FOUNDATION Fieldbus connection

  • FOUNDATION Fieldbus protocol

    FF protocol

  • FOUNDATION Fieldbus protocol standard

    FF design specifications

  • FOUNDATION Fieldbus version

    ITK 4.6

  • FOUNDATION Fieldbus capability

    Basic or LAS

  • FOUNDATION Fieldbus function blocks

    2 analog and 1 PID

Observed authority requirements

  • EMC

    2014/30/EU

  • RoHS

    2011/65/EU

  • EAC

    TR-CU 020/2011

Approvals

  • ATEX 2014/34/EU

    KEMA 02ATEX1318 X

  • IECEx

    BVS 12.0035X

  • FM

    3015609

  • CSA

    1418937

  • INMETRO

    DEKRA 18.0006X

  • NEPSI

    GYJ14.1101X

  • EAC Ex TR-CU 012/2011

    RU C-DK.GB08.V.00410

Your request has been sent successfully. - 0 Items on your list
Product description Quantity
Please specify variant(s) (see data sheet) - or indicate ‘Variant unknown’
Your contact information
Continue browsing
No products added
You have no products on your list...

When browsing product pages on this website, you have the possibility to add products to a list - and send your list to us, thereby requesting a quotation.
You add products to the list by clicking the "GET A QUOTE" button on the individual product pages.

When you have submitted the list, we will get back to you as quickly as possible.

If you don't submit the list, it will be visible to you on the website for 7 days (we will place a so-called' cookie on your pc or tablet).

We hope you will find the functionality useful. Go to the product pages to add products to your list.
Thank you for your request for a quotation.
We will get back to you by phone or email as soon as possible.