HART 7 temperature converter - isolated
- High accuracy, better than 0.05% of span
- Slimline housing of 6 mm
- Excellent EMC performance
- Selectable 60 ms / 60 s response time
- Pre-calibrated temperature ranges selectable via DIP-switches
Application
- The 3113 temperature converter measures a standard Pt100, TC J and K temperature sensor, and provides an isolated active analog current and HART signal output.
- High 3 port isolation provides surge suppression and protects the control system from transients and noise.
- The 3113 can be mounted in the safe area or in Zone 2 / Division 2 areas.
- Approved for marine applications.
Technical characteristics
- Flexibly powered by 24 VDC (±30%) via power rail or connectors.
- A 60 ms fast response time with simultaneous sensor error detection when selected.
- Selectable internal/external CJC.
- Excellent conversion accuracy in all available ranges, better than 0.05% of span.
- Meeting the NAMUR NE21 recommendations, the 3113 provides top measurement performance in harsh EMC environments.
- The device meets the NAMUR NE43 standard defining out of range and sensor error output values.
- A visible green LED indicates operational status of the unit and the input sensor.
- All terminals are protected against overvoltage and polarity error.
- High galvanic isolation of 2.5 kVAC.
- Excellent signal/noise ratio of > 60 dB.
Mounting / installation / programming
- Selectable HART mode with HART 7 revision protocol enables extended device programming.
- Selectable DIP-mode for easy configuration of more than 1000 factory calibrated measurement ranges with HART read only feature.
- The narrow 6 mm housing allows up to 165 units to be mounted per meter of DIN rail, without any air gap between units.
- Wide ambient temperature range of -25...+70°C.
Environmental Conditions
| Operating temperature | -25°C to +70°C |
| Storage temperature | -40°C to +85°C |
| Calibration temperature | 20...28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree | IP20 |
| Installation in | Pollution degree 2 & meas. / overvoltage cat. II |
Mechanical specifications
| Dimensions (HxWxD) | 113 x 6.1 x 115 mm |
| Weight approx. | 70 g |
| DIN rail type | DIN EN 60715/35 mm |
| Wire size | 0.13 x 2.5 mm2 / AWG 26...12 stranded wire |
| Screw terminal torque | 0.5 Nm |
| Vibration | IEC 60068-2-6 |
| 2...25 Hz | ±1.6 mm |
| 25...100 Hz | ±4 g |
Common specifications
Supply |
|
| Supply voltage | 16.8...31.2 VDC |
| Max. required power | 0.70 W |
| Max. power dissipation | 0.70 W |
Isolation voltage |
|
| Isolation voltage, test / working | 2.5 kVAC / 300 VAC (reinforced) |
| Zone 2 / Div. 2 | 250 VAC |
Response time |
|
| HART mode, (0...90%, 100...10%) | 60 ms...60 s, programmable |
| DIP mode, (0...90%, 100...10%) | < 60 ms |
| Accuracy | Better than 0.05% of selected range |
| Signal / noise ratio | Min. 60 dB |
| Long-term stability, better than | ±0.1% of span / year (±0.3% of span / 5 years) |
| Programming | DIP-switches |
| Signal dynamics, input | 23 bit |
| Signal dynamics, output | 18 bit |
| EMC immunity influence | < ±0.5% of span |
| Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
| Incorrect DIP-switch setting identification | 3.5 mA output; LED 0.5 s / 1 Hz |
Input specifications
RTD input |
|
| Temperature range, Pt100 | -200...+850°C |
| Min. measurement range (span) | 10°C |
| Accuracy: the greater of | Better than 0.05% of span or 0.1°C |
| Temperature coefficient: the greater of | 0.02°C/°C or ≤ ±0.01%/°C |
| Sensor current | < 150 µA |
| Sensor cable resistance | < 50 Ω per wire |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
| Sensor error detection | Yes - selectable via DIP-switch |
| Broken sensor detection | > 800 Ω |
| Shorted sensor detection | < 18 Ω |
TC input |
|
| Temperature range, TC J | -100...+1200°C |
| Temperature range, TC K | -180...+1372°C |
| Min. measurement range (span) - TC J & K | 50°C |
| Accuracy: the greater of | Better than 0.05% of span or 0.5°C |
| Temperature coefficient: the greater of | 0.1°C/°C or ≤ ±0.01%/°C |
| Sensor cable resistance | < 5 kΩ per wire |
| Cold junction compensation (CJC): Accuracy @ external Pt100 input | Better than ±0.15°C |
| Cold junction compensation (CJC): Accuracy @ internal CJC | Better than ±2.5°C |
| Internal CJC error detection | Yes |
| External CJC error detection | Yes - selectable via DIP-switch |
| Open Thermocouple detection | Yes - selectable via DIP-switch |
Output specifications
Common output specifications |
|
| Updating time | 10 ms |
Current output |
|
| Signal range | 0...23 mA |
| Programmable signal ranges | 4...20 and 20...4 mA |
| Load (@ current output) | ≤ 600 Ω |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Sensor error indication | 3.5 mA or 23 mA / acc. to NAMUR NE43 or OFF |
| Open output | < 20 V |
| HART protocol revisions | HART 7 |
I.S. / Ex marking
| ATEX | II 3 G Ex nA IIC T4 Gc |
| IECEx | Ex nA IIC T4 Gc |
| FM, US | Cl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, AEx nA IIC T4 |
| FM, CA | Cl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, Ex nA IIC T4 |
Observed authority requirements
| EMC | 2014/30/EU |
| LVD | 2014/35/EU |
| RoHS | 2011/65/EU |
| EAC | TR-CU 020/2011 |
Approvals
| ATEX 2014/34/EU | KEMA 10ATEX0147 X |
| IECEx | KEM 10.0068X |
| FM | FM17US0004X / FM17CA0003X |
| DNV-GL Marine | Stand. f. Certific. No. 2.4 |
| UL | UL 61010-1 |
276USB
Viator USB HART modem
3405
Power connector unit
7423
Ceramic socket for Pt100 sensor
7430B
Pt100 cable sensor
7430C
Pt100 cable sensor
7440
Thermowell for PR 7400 Pt100 sensor
9400_1
Power rail 15 mm profile
9400_2
Power rail 7.5 mm profile
9402
Extra end covers for power rail
9404
Module stop
9410
Power control unit