PROFIBUS PA temperature transmitter
5450B
- PROFIBUS PA profile 4.0
- RTD, TC, potentiometer, linear resistance and bipolar mV input
- Single or true dual inputs with sensor redundancy and drift detection
- Easy Ex i deployment with conformity to FICSO IEC 60079-27
- Accuracy from 0.04°C
- 2.5 kVAC galvanic isolation
Application
- Temperature measurement of a wide range of TC and RTD types.
- Conversion of wide span linear resistance and potentiometer inputs to PROFIBUS.
- Conversion of bipolar mV signals to PROFIBUS.
- Integration into asset management schemes.
- Critical applications requiring superior accuracy and/or sensor redundancy and drift detection.
Technical characteristics
- True dual input transmitter. High density 7-terminal design accepts the widest range of dual input combinations.
- Sensor drift detection - alerts when sensor differential exceeds user-defined limits, for maintenance optimization.
- Variable mapping for process data such as average, differential and min./max. tracking.
- Groundbreaking digital and analog signal accuracy over full input span and ambient conditions.
- Extensive sensor matching including Callendar Van Dusen and custom linearizations.
- Programmable input limits with runtime metering ensure maximum process traceability and sensor out of range protection.
- Meets NAMUR NE21, NE44, NE89 and NE107 compliant diagnostics information.
Programming
- PROFIBUS PA functions and device-specific parameters are configured via fieldbus communication and the associated DD / DTM / GSD device drivers.
Mounting / installation
- For DIN form B sensor head mounting.
- Mounting in zone 0, 1, 2 and zone 21, 22 including M1.
Data sheet & Manuals
Approvals & Certificates
Supplementary documentation
EU Declaration of Conformity
- 5450DoC_EU_100_UK.pdf
- Latest version
EMC data
- 5450_EMC_data.pdf
- Latest version
PR default configurations
- PR_default_configurations.pdf
- Latest version
Safety note
Order form
Environmental Conditions
Operating temperature | -40°C to +80°C |
Storage temperature | -50°C to +85°C |
Calibration temperature | 23...25°C |
Relative humidity | < 99% RH (non-cond.) |
Protection degree (encl./terminal) | IP68 / IP00 |
Mechanical specifications
Dimensions | Ø 44 x 21.45 mm |
Center hole diameter | Ø 6.35 mm / ¼ in |
Weight approx. | 50 g |
Wire size | 1 x 1.5 mm2 stranded wire |
Stripping length | 7 mm |
Screw terminal torque | 0.4 Nm |
Vibration | IEC 60068-2-6 |
2...25 Hz | ±1.6 mm |
25...100 Hz | ±4 g |
Common specifications
Supply |
|
Supply voltage | 9.0...30 VDC |
Supply voltage in FISCO installations | 9.0...17.5 VDC |
Internal power dissipation | < 352 mW |
Isolation voltage |
|
Isolation voltage, test / working | 2.5 kVAC / 42 VAC |
Response time |
|
Response time | < 400 ms |
Programmable damping | 0...60 s |
Write protection | Jumper |
Warm-up time | < 5 min. |
Start-up time | < 15 s |
Programming | PROFIBUS GSD, DD |
Signal / noise ratio | > 60 dB |
Updating time, PROFIBUS | < 100 ms |
Signal dynamics, input | 24 bit |
Accuracy | See manual for details |
EMC immunity influence | < ±0.1% of span |
Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% |
Input specifications
RTD input |
|
RTD type | Pt10...10000, Ni10...10000, Cu5...1000 |
Permissible cable resistance per wire | ≤ 50 Ω |
Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
Sensor current | < 0.15 mA |
Sensor error detection | None, Shorted, Broken, Shorted or Broken |
TC input |
|
Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
Cold junction compensation (CJC) | Constant, internal or external via a Pt100 or Ni100 sensor |
Sensor error detection | None, Shorted, Broken, Shorted or Broken |
Linear resistance input |
|
Measurement range / min. range (span) | 0 Ω...100 kΩ / 25 Ω |
Permissible cable resistance per wire | ≤ 50 Ω |
Sensor current | < 0.15 mA |
Sensor error detection | None, Broken |
Potentiometer input |
|
Potentiometer min....max. | 10 Ω...100 kΩ |
Measurement range / min. range (span) | 0...100% / 10% |
Permissible cable resistance per wire | ≤ 50 Ω |
Sensor current | < 0.15 mA |
Sensor error detection | None, Shorted, Broken, Shorted or Broken |
mV input |
|
Measurement range |
-800...+800 mV (bipolar)
-100...1700 mV |
Min. measurement range (span) | 2.5 mV |
Input resistance | 10 MΩ |
Sensor error detection | None, Broken |
Output specifications
PROFIBUS PA connection |
|
PROFIBUS PA protocol | Profile A&B, ver. 3.0 |
PROFIBUS PA protocol standard | EN 50170 vol. 2 |
PROFIBUS PA address (at delivery) | 126 |
PROFIBUS PA function blocks | 2 analog |
Observed authority requirements
EMC | 2014/30/EU & UK SI 2016/1091 |
ATEX | 2014/34/EU & UK SI 2016/1107 |
RoHS | 2011/65/EU & UK SI 2012/3032 |
Approvals
ATEX | DEKRA 24ATEX0006X |
Need support to select the right product for your application?
Our sales engineers are ready to help specify the right device to meet your needs.
Reach out by phone to get in touch with us straight away – or use the contact form or quick quote function to send your RFQ directly from the website. You’ll get a response within 24 hours on normal business days.
You can expect to get a confirmed delivery date via email within 2 days after we have received your order.
As soon as your package leaves our warehouse, you’ll receive tracking details via email. And if you have any questions along the way we’re just a phone call or email away.
![]() |
State-of-the-art manufacturing Our 8,500 sqm integrated and automated manufacturing campus in Denmark covers the entire value-chain from design and development to manufacturing. It allows us to design and optimize for testing and manufacturing thereby constantly driving quality up and costs down. |
![]() |
Dedicated Presale / Aftersales
|
![]() |
Product reliability |
![]() |
Product deployment
|
![]() |
Smart products
Check out our range of communication interfaces
|