Items 0
5114B
 

5114B

 

Programmable transmitter

 

  • Input for RTD, TC, mV, linear resistance, mA, and V
  • 3-port 3.75 kVAC galvanic isolation
  • Current and voltage output
  • Universal voltage supply
  • 1- and 2-channel versions
  • Loop supply > 17.1 V in Ex / I.S. zone 0
  1. Overview

  2. Technical details

  3. Accessories


Advanced features

  • The 5114 transmitter can be configured using the PReset  software and the Loop Link communications unit.

Application

  • Jumper selectable inputs for current/voltage or temperature.
  • Programmable current (0...100 mA) and voltage (0...250 VDC) inputs.
  • Linearized, electronic temperature measurement.
  • Conversion of linear resistance variation e.g, from solenoids and butterfly valves or linear movements with attached potentiometer.
  • 17.1 VDC loop and 2.5 VDC potentiometer supplies.
  • Automatic 4- / 3-wire or programmable 2-wire cable compensation.
  • Configurable sensor error detection including NAMUR NE43.

Technical characteristics

  • Active or Passive current output and selectable voltage output.
  • Separation of circuits in PELV/SELV installations.
  • I.S. barrier for temperature sensors, potentiometers, and current / voltage signals.
  • I.S. barrier with I.S. power supply for 2-wire transmitters in zone 0, 1, 2, 20, 21 and 2.

Environmental Conditions

Operating temperature -20°C to +60°C
Calibration temperature 20...28°C
Relative humidity < 95% RH (non-cond.)
Protection degree IP20

Mechanical specifications

Dimensions (HxWxD) 109 x 23.5 x 130 mm
Weight approx. 225 g
DIN rail type DIN 46277
Wire size 1 x 2.5 mm2 stranded wire
Screw terminal torque 0.5 Nm
Vibration IEC 60068-2-6
2...13.2 Hz ±1 mm
13.2...100 Hz ±0.7 g

Common specifications

Supply

Supply voltage, universal 21.6...253 VAC, 50...60 Hz or 19.2...300 VDC
Fuse 400 mA SB / 250 VAC
Max. required power 2.1 W / 2.8 W (1 / 2 ch.)

Isolation voltage

Isolation voltage, test / working 3.75 kVAC / 250 VAC
PELV/SELV IEC 61140

Response time

Temperature input, programmable (0...90%, 100...10%) 400 ms...60 s
mA / V input (programmable) 250 ms...60 s

Auxiliary supplies

2-wire supply (pin 44...42 and 54...52) 28...17.1 VDC / 0...20 mA
Programming Loop Link
Signal / noise ratio Min. 60 dB (0...100 kHz)
Accuracy Better than 0.05% of selected range
Updating time 115 ms (temperature input) 75 ms (mA / V / mV input)
Signal dynamics, input 22 bit
Signal dynamics, output 16 bit
Auxiliary voltages: Reference voltage 2.5 VDC ±0.5% / 15 mA
EMC immunity influence < ±0.5% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst < ±1% of span

Input specifications

Common input specifications

Max. offset 50% of selected max. value

RTD input

RTD type Pt100, Ni100, lin. R
Cable resistance per wire 10 Ω (max.)
Sensor current Nom. 0.2 mA
Effect of sensor cable resistance (3-/4-wire) < 0.002 Ω / Ω

TC input

Thermocouple type B, E, J, K, L, N, R, S, T, U, W3, W5, LR
Cold junction compensation (CJC) < ±1.0°C
Sensor error current Nom. 30 μA
Sensor error detection Yes

Current input

Measurement range 0...100 mA
Min. measurement range (span) 4 mA
Input resistance: Supplied unit Nom. 10 Ω + PTC 10 Ω
Input resistance: Non-supplied unit RSHUNT = ∞, VDROP < 6 V

Voltage input

Measurement range 0...250 VDC -150...+150 mV
Min. measurement range (span) 5 mV
Input resistance Nom. 10 MΩ (≤ 2.5 VDC) Nom. 5 MΩ (> 2.5 VDC)
Nom. 10 MΩ (mV input)

Output specifications

Current output

Signal range 0...20 mA
Min. signal range 10 mA
Load (@ current output) ≤ 600 Ω
Load stability ≤ 0.01% of span / 100 Ω
Current limit ≤ 28 mA
Sensor error indication Programmable 0...23 mA
NAMUR NE43 Upscale/Downscale 23 mA / 3.5 mA

Passive 2-wire mA output

Max. load resistance [Ω] (Vsupply-3.5)/0.023 A
Max. external 2-wire supply 29 VDC
Effect of external 2-wire supply voltage variation < 0.005% of span / V

Voltage output

Signal range 0...10 VDC
Min. signal range 500 mV
Load (@ voltage output) ≥ 500 kΩ
2-wire 4...20 mA output: Signal range 4...20 mA
Load stability, 4…20 mA output ≤ 0.01% of span / 100 Ω
of span = of the presently selected range

Observed authority requirements

EMC 2014/30/EU
LVD 2014/35/EU
RoHS 2011/65/EU
EAC TR-CU 020/2011

Approvals

ATEX 2014/34/EU DEMKO 99ATEX124571, II (1) GD [EEx ia] IIC
EAC Ex TR-CU 012/2011 RU C-DK.GB08.V.00410
DNV-GL Marine Stand. f. Certific. No. 2.4
Your request has been sent successfully. - 0 Items on your list
Product description Quantity
Please specify variant(s) (see data sheet) - or indicate ‘Variant unknown’
Your contact information
Continue browsing
No products added
You have no products on your list...

When browsing product pages on this website, you have the possibility to add products to a list - and send your list to us, thereby requesting a quotation.
You add products to the list by clicking the "GET A QUOTE" button on the individual product pages.

When you have submitted the list, we will get back to you as quickly as possible.

If you don't submit the list, it will be visible to you on the website for 7 days (we will place a so-called' cookie on your pc or tablet).

We hope you will find the functionality useful. Go to the product pages to add products to your list.
Thank you for your request for a quotation.