Programmable LED indicator
PR 5714
- 4-digit 14-segment LED display
- Input for mA, V, Ohm, RTD, TC and potentiometer
- 2 relays and analog output
- Universal supply
- Front key programmable
Application
- Display for digital readout of current / voltage / resistance / temperature or potentiometer signals.
- Process control with 2 potential-free relays and / or analog output.
- For local readout in extremely wet atmospheres with a specially designed splash-proof cover.
Technical characteristics
- 4-digit LED indicator with 13.8 mm 14-segment characters. Max. display readout -1999...9999 with programmable decimal point and relay ON / OFF indication.
- All standard operational parameters can be adjusted to any application by way of the front function keys.
- Help texts in eight languages can be selected via a menu item.
- PR 5714 is available fully-configured according to specifications ready for process control and visualization.
- 2.3 kVAC galvanic isolation between input, output and supply.
- In versions with relay outputs the user can minimize the installation test time by activating / deactivating each relay independently of the input signal.
Mounting / installation
- To be mounted in panel front. The included rubber packing must be mounted between the panel cutout hole and the display front to obtain a protection degree of IP65 (type 4X). For extra protection in extreme environments, PR5714 can be delivered with a specially designed splash-proof cover as accessory.
Data sheet & Manuals
Data sheet
- 5714.pdf
- Latest version
Manuals
- 5714V106_UK.pdf
- 5714V105_UK.pdf
- 5714V104_UK.pdf
- 5714V103_UK.pdf
- 5714V102.pdf
- 5714V101.pdf
- 5714V100.pdf
- Previous versions
- Latest version
Manuals in Russian
- 5714V106_RU.pdf
- 5714V105_RU.pdf
- 5714V104_RU.pdf
- 5714V103_RU.pdf
- 5714L102_RU.pdf
- 5714L101_RU.pdf
- Previous versions
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 5714DoC_EU_103_UK.pdf
- 5714DoC_EU_102_UK.pdf
- 5714DoC_EU_101_UK.pdf
- 5714DoC_EU_100_UK.pdf
- Previous versions
- Latest version
PR default configurations
- PR_default_configurations.pdf
- Latest version
Safety note
Safety note
Approvals & Certificates
UL certificate
- 5714_5715_5725_UL.pdf
- Latest version
DNV EU-RO marine certificate
- DNV_EU_RO_4114_4116_4131_5714_5715_5725.pdf
- Latest version
EAC declaration
- EAC_LVD_EMC_Declaration.pdf
- Latest version
Metrology Pattern Approval (MPA), Russia
- MPA_Russia.pdf
- Latest version
View variants
Environmental Conditions
| Operating temperature | -20°C to +60°C |
| Calibration temperature | 20...28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree (mounted in panel) | IP65 / Type 4X |
Mechanical specifications
| Dimensions (HxWxD) | 48 x 96 x 120 mm |
| Cut out dimensions | 44.5 x 91.5 mm |
| Weight approx. | 230 g |
| Wire size, pin 41-46 (max.) | 0.05...1.31 mm2 AWG 30...16 stranded wire |
| Wire size, others, max. | 0.05...3.31 mm2 / AWG 30...12 stranded wire |
| Vibration | IEC 60068-2-6 |
| 2...13.2 Hz | ±1 mm |
| 13.2...100 Hz | ±0.7 g |
Common specifications
Supply |
|
| Supply voltage, universal | 21.6...253 VAC, 50...60 Hz or 19.2...300 VDC |
| Max. required power |
2.5 W (5714A)
3.0 W (5714B/C) 3.5 W (5714D) |
| Internal power dissipation |
2.2 W (5714A)
2.7 W (5714B/C) 3.2 W (5714D) |
Isolation voltage |
|
| Isolation voltage, test / working | 2.3 kVAC / 250 VAC |
Response time |
|
| Temperature input, programmable (0...90%, 100...10%) | 1...60 s |
| mA / V input (programmable) | 0.4...60 s |
Auxiliary supplies |
|
| 2-wire supply (pin 46...45) | > 15 VDC @ 0...20 mA |
| Programming | Front-programmable |
| Signal / noise ratio | Min. 60 dB (0...100 kHz) |
| Accuracy | Better than 0.1% of sel. range |
| EMC immunity influence | < ±0.5% of readout |
Input specifications
RTD input |
|
| RTD type | Pt10/20/50/100/200/250; Pt300/400/500/1000; Ni50/100/120/1000; Cu10/20/50/100 |
| Cable resistance per wire | 50 Ω (max.) |
| Sensor current | Nom. 0.2 mA |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
Linear resistance input |
|
| Linear resistance min....max. | 0 Ω...10000 Ω |
Potentiometer input |
|
| Potentiometer min....max. | 10 Ω...100 kΩ |
TC input |
|
| Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
| CJC via int. mounted sensor | ±(2.0°C + 0.4°C * Δt) |
| Sensor error detection | Yes |
| Sensor error current: When detecting / else | Nom. 2 μA / 0 μA |
Current input |
|
| Measurement range | 0...23 mA |
| Programmable measurement ranges | 0...20 and 4...20 mA |
| Input resistance | Nom. 20 Ω + PTC 25 Ω |
| Sensor error detection | Loop break 4...20 mA |
Voltage input |
|
| Measurement range | 0...12 VDC |
| Programmable measurement ranges | 0/0.2...1; 0/2...10 VDC |
| Input resistance | Nom. 10 MΩ |
Output specifications
Display |
|
| Display readout | -1999...9999 (4 digits) |
| Decimal point | Programmable |
| Digit height | 13.8 mm |
| Display updating. | 2.2 times / s |
| Input outside input range is indicated by | Explanatory text |
Current output |
|
| Signal range | 0...23 mA |
| Programmable signal ranges | 0...20/4...20/20...0/20...4 mA |
| Load (@ current output) | ≤ 800 Ω |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Sensor error indication | 0 / 3.5 / 23 mA / none |
| NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
| Output limitation, on 4...20 and 20...4 mA signals | 3.8...20.5 mA |
| Output limitation, on 0...20 and 20...0 mA signals | 0...20.5 mA |
| Current limit | ≤ 28 mA |
Relay output |
|
| Relay functions | Setpoint |
| Hysteresis | 0...100% |
| ON and OFF delay | 0...3600 s |
| Sensor error reaction | Break / Make / Hold |
| Max. voltage | 250 VAC / VDC |
| Max. current | 2 A |
| Max. AC power | 500 VA |
| Max. DC current, resistive load > 30 VDC | See manual for details |
Observed authority requirements
| EMC | 2014/30/EU |
| LVD | 2014/35/EU |
| RoHS | 2011/65/EU |
| EAC | TR-CU 020/2011 |
| EAC LVD | TR-CU 004/2011 |
Approvals
| c UL us, UL 508 | E248256 |
| EU RO MR Type Approval | MRA000000Z |