2-wire programmable RTD transmitter
PR 5332A
- RTD or Ohm input
- Accuracy: Better than 0.05% of selected range
- Programmable sensor error value
- For DIN form B sensor head mounting
Application
- Linearized temperature measurement with Pt100...Pt1000 or Ni100...Ni1000 sensor.
- Conversion of linear resistance variation to a standard analog current signal, for instance from valves or Ohmic level sensors.
Technical characteristics
- Within a few seconds the user can program PR5332A to measure temperatures within all ranges defined by the norms.
- Dedicated programmable non-isolated 4-wire RTD transmitter.
- RTD and resistance inputs have cable compensation for 2-, 3- and 4-wire connection.
- Continuous check of vital stored data for safety reasons.
Mounting / installation
- For DIN form B sensor head or DIN rail mounting with the PR fitting type 8421.
Data sheet & Manuals
Data sheet
- 5332A.pdf
- Latest version
Manuals
- 5332V107_UK.pdf
- 5332V106_UK.pdf
- 5332V105_UK.pdf
- 5332V104_UK.pdf
- 5332V103_UK.pdf
- 5332V102_UK.pdf
- 5332V101_UK.pdf
- 5332V100_UK.pdf
- Previous versions
- Latest version
Manuals in Russian
- 5332V107_RU.pdf
- 5332V106_RU.pdf
- 5332V105_RU.pdf
- 5332V104_RU.pdf
- 5332V103_RU.pdf
- 5332V102_RU.pdf
- 5332V101_RU.pdf
- 5332V100_RU.pdf
- Previous versions
- Latest version
Approvals & Certificates
ATEX certificate
- 5332A_20ATEX0096X_issue1.pdf
- 5332A_20ATEX0096X_issue0.pdf
- 5332A_10ATEX0002X_issue3.pdf
- Previous versions
- Latest version
IECEx certificate
- 5332AIEC_200059X_issue1.pdf
- 5332AIEC_200059X_issue0.pdf
- 5332AIEC_130035X_issue1.pdf
- Previous versions
- Latest version
CSA certificate
- 533x_633x_CSA.pdf
- Latest version
INMETRO certificate
- INMETRO_5331_5332_5334_6331_6334_230009X_issue1.pdf
- INMETRO_5331_5332_5334_6331_6334_230009X_issue0.pdf
- Previous versions
- Latest version
EAC declaration
- EAC_EMC_declaration_Ex_3000_5000_6000_9000.pdf
- Latest version
EAC declaration
- EAC_Ex.pdf
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 5332DoC_EU_106_UK.pdf
- 5332DoC_EU_105_UK.pdf
- 5332DoC_EU_104_UK.pdf
- 5332DoC_EU_103_UK.pdf
- Previous versions
- Latest version
CCC statement
- 5000_6000_series_CCC_statement.pdf
- Latest version
EAC LVD declaration
- 5331_5332_5334_EMC_data.pdf
- Latest version
PR default configurations
- PR_default_configurations.pdf
- Latest version
Safety note
Safety note
- SN5332-107-2624.pdf
- SN5332-106-2605.pdf
- SN5332-105-2422.pdf
- SN5332-104-2245.pdf
- SN5332-103-2202.pdf
- SN5332-102-2122.pdf
- SN5332-101-2007.pdf
- SN5332-100-1848.pdf
- Previous versions
- Latest version
View variants
Environmental Conditions
| Operating temperature | -40°C to +85°C |
| Calibration temperature | 20...28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree (encl./terminal) | IP68 / IP00 |
Mechanical specifications
| Dimensions | Ø 44 x 20.2 mm |
| Weight approx. | 50 g |
| Wire size | 1 x 1.5 mm2 stranded wire |
| Screw terminal torque | 0.4 Nm |
Common specifications
Supply |
|
| Supply voltage | 7.2...35 VDC |
| Internal power dissipation | 25 mW…0.8 W |
Response time |
|
| Response time (programmable) | 1...60 s |
| Voltage drop | 7.2 VDC |
| Warm-up time | 5 min. (1) |
| Power on to stable output | 4.5 s |
| Programming | PR 5909 Loop Link communications interface |
| Signal / noise ratio | > 60 dB |
| EEprom error check | < 3.5 s |
| Accuracy | Better than 0.05% of selected range |
| Signal dynamics, input | 20 bit |
| Signal dynamics, output | 16 bit |
| Effect of supply voltage change | < 0.005% of span / VDC |
| EMC immunity influence | < ±0.5% of span |
| Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
Input specifications
Common input specifications |
|
| Max. offset | 50% of selected max. value |
RTD input |
|
| RTD type | Pt100…1000, Ni100...1000, lin. R |
| Cable resistance per wire | 5 Ω (max.) |
| Sensor current | Nom. 0.2 mA |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
| Sensor error detection | Yes |
Linear resistance input |
|
| Linear resistance min....max. | 0 Ω...5000 Ω |
Output specifications
Common output specifications |
|
| Updating time | 440 ms |
Current output |
|
| Signal range | 4…20 mA |
| Min. signal range | 16 mA |
| Load (@ current output) | ≤ (Vsupply - 7.2) / 0.023 [Ω] |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Sensor error indication | Programmable 3.5…23 mA |
| NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
| of span | = of the presently selected range |
I.S. / Ex marking
| ATEX | II 3 G Ex nA [ic] IIC T6...T4 Gc, II 3 G Ex ec [ic] IIC T6...T4 Gc, II 3 G Ex ic IIC T6...T4 Gc, II 3 D Ex ic IIIC Dc |
| IECEx | Ex nA [ic] IIC T6…T4 Gc, Ex ec [ic] IIC T6…T4 Gc, Ex ic IIC T6…T4 Gc, Ex ic IIIC Dc |
| CSA | Cl. I, Div. 2, Gp. A, B, C, D T6...T4, Ex nA[ic] IIC T6…T4 Gc |
| INMETRO | Ex ec [ic] IIC T6...T4 Gc, Ex ic IIC T6...T4 Gc, Ex ic IIIC Dc |
Observed authority requirements
| EMC | 2014/30/EU |
| ATEX | 2014/34/EU |
| RoHS | 2011/65/EU |
| EAC | TR-CU 020/2011 |
| EAC Ex | TR-CU 012/2011 |
Approvals
| ATEX | DEKRA 20ATEX0096X |
| IECEx | DEK 20.0059X |
| CSA | 1125003 |
| INMETRO | DEKRA 23.0009X |
| EAC Ex | EAEU KZ 7500361.01.01.08756 |