2-wire transmitter with HART protocol
5335A
- RTD, TC, Ohm, or mV input
- Extremely high measurement accuracy
- HART 5 protocol
- Programmable sensor error value
- For DIN form B sensor head mounting
Application
- Linearized temperature measurement with Pt100...Pt1000, Ni100...Ni1000, or TC sensor.
- Difference or average temperature measurement of 2 resistance or TC sensors.
- Conversion of linear resistance variation to a standard analog current signal, for instance from valves or Ohmic level sensors.
- Amplification of a bipolar mV signal to a standard 4...20 mA current signal.
- Connection of up to 15 transmitters to a digital 2-wire signal with HART communication.
Technical characteristics
- Within a few seconds the user can program PR5335A to measure temperatures within all ranges defined by the norms.
- The RTD and resistance inputs have cable compensation for 2-, 3- and 4-wire connection.
- The 5335A has been designed according to strict safety requirements and is therefore suitable for application in SIL installations.
- Continuous check of vital stored data for safety reasons.
- Sensor error detection according to the guidelines in NAMUR NE89.
Mounting / installation
- For DIN form B sensor head or DIN rail mounting with the PR fitting type 8421.
Data sheet & Manuals
Data sheet
- 5335A.pdf
- Latest version
Manuals
- 5335V120_UK.pdf
- 5335V119_UK.pdf
- 5335V118_UK.pdf
- 5335V117_UK.pdf
- 5335V116_UK.pdf
- 5335V115_UK.pdf
- 5335V114_UK.pdf
- 5335V113.pdf
- 5335V112.pdf
- 5335V111.pdf
- 5335V110.pdf
- 5335V109.pdf
- 5335V108.pdf
- 5335V107.pdf
- Previous versions
- Latest version
Manuals in Russian
- 5335V120_RU.pdf
- 5335V119_RU.pdf
- 5335V118_RU.pdf
- 5335V117_RU.pdf
- 5335V116_RU.pdf
- 5335V115_RU.pdf
- 5335L112_RU.pdf
- 5335L111_RU.pdf
- 5335L110_RU.pdf
- Previous versions
- Latest version
Approvals & Certificates
ATEX certificate
- 5335A_20ATEX0109X_issue0.pdf
- 5335A_03ATEX1508X_issue06.pdf
- 5335A_03ATEX1508X_issue05.pdf
- 5335A_03ATEX1508X_issue04.pdf
- 5335A_03ATEX1508X_issue03.pdf
- 5335A_03ATEX1508X_issue02.pdf
- Previous versions
- Latest version
IECEx certificate
- 5335AIEC_200063X_issue0.pdf
- 5335AIEC_100083X_issue04.pdf
- 5335AIEC_100083X_issue03.pdf
- Previous versions
- Latest version
CSA certificate
- 533x_633x_CSA.pdf
- Latest version
INMETRO certificate
- INMETRO_5335_5337_180002x_issue1.pdf
- INMETRO_5335_5337_180002x_issue0.pdf
- Previous versions
- Latest version
DNV marine certificate
- DNV_2231_41xx_51xx_53xx.pdf
- Latest version
EAC declaration
- EAC-Declaration.pdf
- Latest version
EAC Ex certificate
- EAC_Ex.pdf
- Latest version
Metrology Pattern Approval (MPA), Belarus
- MPA_Belarus.pdf
- Latest version
Metrology Pattern Approval (MPA), Kazakhstan
- MPA_Kazakhstan.pdf
- Latest version
Metrology Pattern Approval (MPA), Russia
- MPA_Russia.pdf
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 5335DoC_EU_106_UK.pdf
- 5335DoC_EU_105_UK.pdf
- 5335DoC_EU_104_UK.pdf
- 5335DoC_EU_103_UK.pdf
- 5335DoC_EU_102_UK.pdf
- 5335DoC_EU_101_UK.pdf
- 5335DoC_EU_100_UK.pdf
- 5335DoC_EU_02000.pdf
- 5335DoC_EU_01997.pdf
- Previous versions
- Latest version
CCC statement
- 5000_6000_series_CCC_statement.pdf
- Latest version
FMEDA report
- PR_5335_R026_V3R2.pdf
- Latest version
SIL declaration
- 5335_6335_5337_6337_7501_SIL1_declaration_Potmeter_V2.pdf
- 5335_6335_5337_6337_7501_SIL2_declaration_V2.pdf
- Previous versions
- Latest version
HART certificate
- 7501_HART_5_certificate.pdf
- Latest version
Safety note
Safety note
Environmental Conditions
Operating temperature | -40°C to +85°C |
Calibration temperature | 20...28°C |
Relative humidity | < 95% RH (non-cond.) |
Protection degree (encl./terminal) | IP68 / IP00 |
Mechanical specifications
Dimensions | Ø 44 x 20.2 mm |
Weight approx. | 50 g |
Wire size | 1 x 1.5 mm2 stranded wire |
Screw terminal torque | 0.4 Nm |
Vibration | IEC 60068-2-6 |
2...25 Hz | ±1.6 mm |
25...100 Hz | ±4 g |
Common specifications
Supply |
|
Supply voltage | 8.0...35 VDC |
Isolation voltage |
|
Isolation voltage, test / working | 1.5 kVAC / 50 VAC |
Response time |
|
Response time (programmable) | 1...60 s |
Warm-up time | 30 s |
Programming | Loop Link & HART |
Signal / noise ratio | Min. 60 dB |
Accuracy | Better than 0.05% of selected range |
Signal dynamics, input | 22 bit |
Signal dynamics, output | 16 bit |
Effect of supply voltage change | < 0.005% of span / VDC |
EMC immunity influence | < ±0.1% of span |
Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
Input specifications
Common input specifications |
|
Max. offset | 50% of selected max. value |
RTD input |
|
RTD type | Pt100, Ni100, lin. R |
Cable resistance per wire | 5 Ω (up to 50 Ω per wire is possible with reduced measurement accuracy) |
Sensor current | Nom. 0.2 mA |
Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
Sensor error detection | Yes |
TC input |
|
Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5 |
Cold junction compensation (CJC) | < ±1.0°C |
Sensor error detection | Yes |
Sensor error current: When detecting / else | Nom. 33 μA / 0 μA |
Voltage input |
|
Measurement range | -800...+800 mV |
Min. measurement range (span) | 2.5 mV |
Input resistance | 10 MΩ |
Output specifications
Current output |
|
Signal range | 4…20 mA |
Min. signal range | 16 mA |
Load (@ current output) | ≤ (Vsupply - 8) / 0.023 [Ω] |
Load stability | ≤ 0.01% of span / 100 Ω |
Sensor error indication | Programmable 3.5…23 mA |
NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
of span | = of the presently selected range |
I.S. / Ex marking
ATEX | II 3 G Ex nA [ic] IIC T6...T4 Gc, II 3 G Ex ex [ic] IIC T6...T4 Gc, II 3 D Ex ic IIIC Dc |
IECEx | Ex nA [ic] IIC T6…T4 Gc, Ex ec [ic] IIC T6…T4 Gc, Ex ic IIC T6…T4 Gc, Ex ic IIIC Dc |
CSA | Cl. I, Div. 2, Gp. A, B, C, D T6...T4, Ex nA[ic] IIC T6…T4 Gc |
INMETRO | Ex nA [ic] IIC T6...T4 Gc, Ex ic IIC T6...T4 Gc, Ex ic IIIC Dc |
Observed authority requirements
EMC | 2014/30/EU |
ATEX | 2014/34/EU |
RoHS | 2011/65/EU |
EAC | TR-CU 020/2011 |
EAC Ex | TR-CU 012/2011 |
Approvals
ATEX | DEKRA 20ATEX0109X |
IECEx | DEK 20.0063X |
CSA | 1125003 |
INMETRO | DEKRA 18.0002X |
DNV Marine | TAA0000101 |
SIL | Hardware assessed for use in SIL applications |

7400
Pt100 temperature sensor
7401
Instrument housing
7430B
Pt100 cable sensor
7430C
Pt100 cable sensor
7440
Thermowell for PR 7400 Pt100 sensor
277
USB HART modem
278
Bluetooth Low Energy HART modem
5909
Loop Link communications interface
7009
10-turn potentiometer, 200 Ω
7011
Dial for 10-turn potentiometer
7012
1-turn potentiometer, 1 kΩ
7020
Knob for 1-turn potentiometer
7020A
Knob for 10-turn potentiometer
7028
10-turn potentiometer, 2 kΩ