2-wire programmable transmitter
6331B
- RTD, TC, Ohm, or mV input
- Extremely high measurement accuracy
- Galvanic isolation
- Can be installed in Ex zone 0
- 1- or 2-channel version
Application
- Linearized temperature measurement with Pt100...Pt1000, Ni100...Ni1000, or TC sensor.
- Conversion of linear resistance variation to a standard analog current signal, for instance from valves or Ohmic level sensors.
- Amplification of a bipolar mV signal to a standard 4...20 mA current signal.
Technical characteristics
- Within a few seconds the user can program PR6331B to measure temperatures within all ranges defined by the norms.
- The RTD and resistance inputs have cable compensation for 2-, 3- and 4-wire connection.
- A limit can be programmed on the output signal.
- Continuous check of vital stored data for safety reasons.
Mounting / installation
- Mounted vertically or horizontally on a DIN rail. Using the 2-channel version, up to 84 channels can be mounted per meter.
- The 6331B can be mounted in zone 0, 1, 2 and zone 20, 21, 22 including M1 / Class I, Division 1, Groups A, B, C, D.
Data sheet & Manuals
Data sheet
- 6331B.pdf
- Latest version
Manuals
- 6331V107_UK.pdf
- 6331V106_UK.pdf
- 6331V105_UK.pdf
- 6331V104.pdf
- 6331V103.pdf
- Previous versions
- Latest version
Manuals in Russian
- 6331V107_RU.pdf
- 6331V106_RU.pdf
- 6331V105_RU.pdf
- 6331L104_RU.pdf
- Previous versions
- Latest version
Approvals & Certificates
ATEX certificate
- 6331B_issue05.pdf
- 6331B_issue04.pdf
- 6331B_issue03.pdf
- 6331B_issue02.pdf
- Previous versions
- Latest version
IECEx certificate
- 6331BIEC_issue02.pdf
- 6331BIEC_issue01.pdf
- 6331BIEC_issue0.pdf
- Previous versions
- Latest version
CSA certificate
- 533x_633x_CSA.pdf
- Latest version
FM certificate
- 53xx_633x_FM17US0013X.pdf
- 53xx_633x_FM_2D5A7.pdf
- Previous versions
- Latest version
EAC declaration
- EAC-Declaration.pdf
- Latest version
EAC Ex certificate
- EAC_Ex.pdf
- Latest version
Metrology Pattern Approval (MPA), Belarus
- MPA_Belarus.pdf
- Latest version
Metrology Pattern Approval (MPA), Kazakhstan
- MPA_Kazakhstan.pdf
- Latest version
Metrology Pattern Approval (MPA), Russia
- MPA_Russia.pdf
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 6331DoC_EU_102_UK.pdf
- 6331DoC_EU_101_UK.pdf
- 6331DoC_EU_100_UK.pdf
- Previous versions
- Latest version
CCC statement
- 5000_6000_series_CCC_statement.pdf
- Latest version
Safety note
Safety note
- SN6300-1_106_2010.pdf
- SN6300-1_105_1749.pdf
- SN6300-1_104_1650.pdf
- SN6300-1_103_1614.pdf
- SN6300_102_1506.pdf
- Previous versions
- Latest version
Environmental Conditions
Operating temperature | -40°C to +85°C |
Storage temperature | -40°C to +85°C |
Calibration temperature | 20...28°C |
Relative humidity | < 95% RH (non-cond.) |
Protection degree | IP20 |
Mechanical specifications
Dimensions (HxWxD) | 109 x 23.5 x 104 mm |
Weight (1 / 2 channels) | 145 / 185 g |
DIN rail type | DIN EN 60715/35 mm |
Wire size | 0.13...2.08 mm2 AWG 26...14 stranded wire |
Screw terminal torque | 0.5 Nm |
Common specifications
Supply |
|
Supply voltage | 7.2...30 VDC |
Max. required power | ≤ 0.7 W/≤ 1.4 W (1 ch./2 ch.) |
Power dissipation, per channel | 0.17...0.7 W |
Isolation voltage |
|
Isolation voltage, test / working | 1.5 kVAC / 50 VAC |
Response time |
|
Response time (programmable) | 1...60 s |
Voltage drop | 7.2 VDC |
Warm-up time | 5 min. |
Programming | Loop Link |
Signal / noise ratio | Min. 60 dB |
Accuracy | Better than 0.05% of selected range |
EEprom error check | < 3.5 s |
Signal dynamics, input | 20 bit |
Signal dynamics, output | 16 bit |
Effect of supply voltage change | < 0.005% of span / VDC |
EMC immunity influence | < ±0.5% of span |
Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
Input specifications
Common input specifications |
|
Max. offset | 50% of selected max. value |
RTD input |
|
RTD type | Pt100, Ni100, lin. R |
Cable resistance per wire | 5 Ω (max.) |
Sensor current | Nom. 0.2 mA |
Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
Sensor error detection | Yes |
TC input |
|
Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
Cold junction compensation (CJC) | < ±1.0°C |
Sensor error detection | Yes |
Sensor error current: When detecting / else | Nom. 33 μA / 0 μA |
Linear resistance input |
|
Linear resistance min....max. | 0 Ω...5000 Ω |
Voltage input |
|
Measurement range | -12...800 mV |
Min. measurement range (span) | 5 mV |
Input resistance | 10 MΩ |
Output specifications
Current output |
|
Signal range | 4…20 mA |
Min. signal range | 16 mA |
Load (@ current output) | ≤ (Vsupply - 7.2) / 0.023 [Ω] |
Load stability | ≤ 0.01% of span / 100 Ω |
Sensor error indication | Programmable 3.5…23 mA |
NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
Common output specifications |
|
Updating time | 440 ms |
of span | = of the presently selected range |
Observed authority requirements
EMC | 2014/30/EU |
ATEX | 2014/34/EU |
RoHS | 2011/65/EU |
EAC | TR-CU 020/2011 |
EAC Ex | TR-CU 012/2011 |
Approvals
ATEX | KEMA 06ATEX0115 X |
IECEx | DEK 14.0047 X |
CSA | 1125003 |
c FM us | FM17US0013X |
EAC Ex | RU C-DK.HA65.B.00355/19 |

7423
Ceramic socket for Pt100 sensor
7430B
Pt100 cable sensor
7430C
Pt100 cable sensor
7440
Thermowell for PR 7400 Pt100 sensor
5909
Loop Link communications interface
5910Ex
CJC connector
5913Ex
CJC connector
7007
2-digit digital potentiometer
7008
3-digit digital potentiometer
7009
10-turn potentiometer, 200 Ω
7011
Dial for 10-turn potentiometer
7012
1-turn potentiometer, 1 kΩ
7020
Knob for 1-turn potentiometer
7020A
Knob for 10-turn potentiometer