Communication enabler
4511
- Programming display for system 4000 and 9000 devices
- Modbus RTU protocol interface over RS-485
- Monitor process value from the built-in display
- High 2.5 kV isolation to host unit
- Shielded RJ45 connector on top
Application
- The 4511 detachable display adds Modbus RTU RS-485 serial communications to all current and future 4000/9000 units.
- The unit converts a wide array of sensors and analog device signals measured by the system 4000 like uni- and bipolar mA and voltage signals, potentiometer, Lin. R, RTD and TC, to a Modbus communication line signal.
- When mounted on a system 9000 device any signal coming from or going to I.S. classified area, like AI, AO, DI and DO signals, can be converted to a Modbus network.
- All individual unit operating parameters can easily and quickly be configured by using the Modbus communication or by using the front display menu.
- The easily readable 4511 display can be used to read the process signal, simulate the output signal, indicate sensor errors and internal device errors.
Technical characteristics
- 4511 has full 4501 functionality for unit programming, process signal monitoring and diagnostics handling.
- Modbus RTU protocol is supported using a serial RS-485 communication wiring.
- Multidrop half-duplex connection via shielded RJ45 connector.
- High safe galvanic isolation of 2.5 kVAC between the serial wiring and the connected system 4000/9000 units.
- Modbus parameters such as address, baud rate, stop bit(s), and parity bit are configured from the 4511 display, which also stores parameters.
Mounting / installation / programming
- Mounting in Zone 2 / Div 2.
- The 4511 can be moved from one device to another. The individual system 4000/9000 unit configuration of the first device can be saved and downloaded to subsequent devices.
- Programmed parameters can be protected by a userdefined password.
- When mounted on devices that are installed upside down, a menu item allows the display on the 4511 to be rotated 180o and the up/down buttons to switch function.
Data sheet & Manuals
Data sheet
- 4511.pdf
- Latest version
Manuals
- 4511V102_UK.pdf
- 4511V101_UK.pdf
- 4511V100_UK.pdf
- Previous versions
- Latest version
Manuals in Russian
- 4511V102_RU.pdf
- 4511V101_RU.pdf
- 4511V100_RU.pdf
- Previous versions
- Latest version
Approvals & Certificates
ATEX certificate
- 4511_issue3.pdf
- Latest version
IECEx certificate
- 4511IEC_issue2.pdf
- Latest version
FM certificate
- FM_4511.pdf
- Latest version
UL certificate
- 4511_UL.pdf
- Latest version
DNV marine certificate
- DNV_9xxx_4511.pdf
- Latest version
EAC declaration
- EAC-Declaration.pdf
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 4511DoC_EU_103_UK.pdf
- 4511DoC_EU_102_UK.pdf
- 4511DoC_EU_101_UK.pdf
- 4511DoC_EU_100_UK.pdf
- Previous versions
- Latest version
Safety note
Safety note
- SN4511_104_2124.pdf
- SN4511_102_1811.pdf
- SN4511_101_1613.pdf
- SN4511_100_1350.pdf
- Previous versions
- Latest version
Environmental Conditions
Operating temperature | -20°C to +60°C |
Storage temperature | -20°C to +85°C |
Relative humidity | < 95% RH (non-cond.) |
Protection degree | IP20 |
Installation in | Pollution degree 2 & meas. / overvoltage cat. II |
Mechanical specifications
Dimensions (HxWxD) | 73.2 x 23.3 x 26.5 mm |
Dimensions (HxWxD) w/ 4000/9000 unit | 109 x 23.5 x 131 mm |
Weight approx. | 30 g |
Connection | RJ45 - shielded |
Common specifications
Supply |
|
Max. required power | 0.15 W |
Isolation voltage |
|
Isolation voltage, test / working | 2.5 kVAC / 250 VAC reinforced isolation |
Response time |
|
Response time | < 20 ms |
Signal / noise ratio | > 60 dB |
Update rate | > 50 Hz |
Extended EMC immunity: NAMUR NE21, A criterion, burst | No loss of communication |
Signal type | RS-485 half duplex |
Serial protocol | Modbus RTU |
Modbus mode | RTU - slave |
Devices on an RS485 line | Up to 32 (w/o a repeater) |
Data rates, baud | 2400, 4800, 9600, 19200, 38400, 57600, 115200 |
Automatic baudrate detection | Yes - can be configured ON or OFF |
Parity | Even, Odd, None |
Stop bit(s) | 1 or 2 |
Digital addressing | 1...247 |
Response delay | 0...1000 ms |
Observed authority requirements
EMC | 2014/30/EU |
LVD | 2014/35/EU |
RoHS | 2011/65/EU |
EAC | TR-CU 020/2011 |
Approvals
ATEX | DEKRA 13ATEX0098 X |
IECEx | DEK 13.0026 X |
FM | 0003049132-C |
UL | UL 61010-1 |
DNV Marine | TAA00000JD |