2-wire programmable transmitter
PR 5334A
- TC input
- High measurement accuracy
- Galvanic isolation
- Programmable sensor error value
- For DIN form B sensor head mounting
Application
- Linearized temperature measurement with TC sensor.
- Amplification of bipolar mV signals to a 4...20 mA signal, optionally linearized according to a defined linearization function.
Technical characteristics
- Within a few seconds the user can program PR5334A to measure temperatures within all TC ranges defined by the norms.
- Cold junction compensation (CJC) with a built-in temperature sensor.
- Continuous check of vital stored data for safety reasons.
Mounting / installation
- For DIN form B sensor head or DIN rail mounting with the PR fitting type 8421.
Data sheet & Manuals
Data sheet
- 5334A.pdf
- Latest version
Manuals
- 5334V115_UK.pdf
- 5334V114_UK.pdf
- 5334V113_UK.pdf
- 5334V112_UK.pdf
- 5334V111_UK.pdf
- 5334V110_UK.pdf
- 5334V109_UK.pdf
- 5334V108_UK.pdf
- 5334V107_UK.pdf
- 5334V106.pdf
- 5334V105.pdf
- 5334V103.pdf
- Previous versions
- Latest version
Manuals in Russian
- 5334V115_RU.pdf
- 5334V114_RU.pdf
- 5334V113_RU.pdf
- 5334V112_RU.pdf
- 5334V111_RU.pdf
- 5334V110_RU.pdf
- 5334V109_RU.pdf
- 5334V108_RU.pdf
- 5334V107_RU.pdf
- 5334L106_RU.pdf
- Previous versions
- Latest version
Manuals in Portuguese (Brazil)
- 5334V115_BR.pdf
- 5334V114_BR.pdf
- 5334V113_BR.pdf
- 5334V112_BR.pdf
- 5334V111_BR.pdf
- 5334V110_BR.pdf
- 5334V109_BR.pdf
- 5334V108_BR.pdf
- Previous versions
- Latest version
Approvals & Certificates
ATEX certificate
- 5334A_20ATEX0096X_issue0.pdf
- 5334A_10ATEX0002X_issue03.pdf
- 5334A_10ATEX0002X_issue02.pdf
- 5334A_10ATEX0002X_issue01.pdf
- Previous versions
- Latest version
IECEx certificate
- 5334AIEC_200059X_issue0.pdf
- 5334AIEC_130035X_issue1.pdf
- 5334AIEC_130035X_issue0.pdf
- Previous versions
- Latest version
INMETRO certificate
- INMETRO_5331_5332_5334_6331_6334_230009X_issue0.pdf
- INMETRO_5331_5332_5334_160013X_issue1.pdf
- INMETRO_5331_5334_160013X_issue0.pdf
- Previous versions
- Latest version
DNV marine certificate
- DNV_2231_41xx_51xx_53xx.pdf
- Latest version
EAC declaration
- EAC_EMC_declaration_Ex_3000_5000_6000_9000.pdf
- Latest version
EAC Ex certificate
- EAC_Ex.pdf
- Latest version
Metrology Pattern Approval (MPA), Belarus
- MPA_Belarus.pdf
- Latest version
Metrology Pattern Approval (MPA), Kazakhstan
- MPA_Kazakhstan.pdf
- Latest version
Metrology Pattern Approval (MPA), Russia
- MPA_Russia.pdf
- Latest version
Supplementary documentation
EU Declaration of Conformity
- 5334DoC_EU_106_UK.pdf
- 5334DoC_EU_105_UK.pdf
- 5334DoC_EU_104_UK.pdf
- 5334DoC_EU_103_UK.pdf
- 5334DoC_EU_102_UK.pdf
- 5334DoC_EU_101_UK.pdf
- 5334DoC_EU_100_UK.pdf
- Previous versions
- Latest version
UKCA Declaration of Conformity
- 5334DoC_UKCA_100_UK.pdf
- Latest version
CCC statement
- 5000_6000_series_CCC_statement.pdf
- Latest version
EMC data
- 5331_5332_5334_EMC_data.pdf
- Latest version
PR default configurations
- PR_default_configurations.pdf
- Latest version
Safety note
Safety note
Order form
Environmental Conditions
Operating temperature | -40°C to +85°C |
Calibration temperature | 20...28°C |
Relative humidity | < 95% RH (non-cond.) |
Protection degree (encl./terminal) | IP68 / IP00 |
Mechanical specifications
Dimensions | Ø 44 x 20.2 mm |
Weight approx. | 50 g |
Wire size | 1 x 1.5 mm2 stranded wire |
Screw terminal torque | 0.4 Nm |
Vibration | IEC 60068-2-6 |
2...25 Hz | ±1.6 mm |
25...100 Hz | ±4 g |
Common specifications
Supply |
|
Supply voltage | 7.2...35 VDC |
Internal power dissipation | 25 mW…0.8 W |
Isolation voltage |
|
Isolation voltage, test / working | 1.5 kVAC / 50 VAC |
Response time |
|
Response time (programmable) | 1...60 s |
Voltage drop | 7.2 VDC |
Warm-up time | 5 min. |
Power on to stable output | 4.5 s |
Programming | PR 5909 Loop Link communications interface |
Signal / noise ratio | Min. 60 dB |
EEprom error check | < 3.5 s |
Accuracy | Better than 0.05% of selected range |
Signal dynamics, input | 18 bit |
Signal dynamics, output | 16 bit |
Effect of supply voltage change | < 0.005% of span / VDC |
EMC immunity influence | < ±0.5% of span |
Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
Input specifications
Common input specifications |
|
Max. offset | 50% of selected max. value |
TC input |
|
Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
Cold junction compensation (CJC) | < ±1.0°C |
Sensor error detection | Yes |
Sensor error current: When detecting / else | Nom. 33 μA / 0 μA |
Voltage input |
|
Measurement range | -12...150 mV |
Min. measurement range (span) | 5 mV |
Input resistance | 10 MΩ |
Output specifications
Current output |
|
Signal range | 4…20 mA |
Min. signal range | 16 mA |
Load (@ current output) | ≤ (Vsupply - 7.2) / 0.023 [Ω] |
Load stability | ≤ 0.01% of span / 100 Ω |
Sensor error indication | Programmable 3.5…23 mA |
NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
Common output specifications |
|
Updating time | 440 ms |
of span | = of the presently selected range |
I.S. / Ex marking
ATEX | II 3 G Ex nA [ic] IIC T6...T4 Gc, II 3 G Ex ec [ic] IIC T6...T4 Gc, II 3 G Ex ic IIC T6...T4 Gc, II 3 D Ex ic IIIC Dc |
IECEx | Ex nA [ic] IIC T6…T4 Gc, Ex ec [ic] IIC T6…T4 Gc, Ex ic IIC T6…T4 Gc, Ex ic IIIC Dc |
INMETRO | Ex ec [ic] IIC T6...T4 Gc, Ex ic IIC T6...T4 Gc, Ex ic IIIC Dc |
Observed authority requirements
EMC | 2014/30/EU & UK SI 2016/1091 |
ATEX | 2014/34/EU & UK SI 2016/1107 |
RoHS | 2011/65/EU & UK SI 2012/3032 |
EAC | TR-CU 020/2011 |
EAC Ex | TR-CU 012/2011 |
Approvals
ATEX | DEKRA 20ATEX0096X |
IECEx | DEK 20.0059X |
INMETRO | DEKRA 23.0009X |
EAC Ex | EAEU KZ 7500361.01.01.08756 |
DNV Marine | TAA0000101 |
Do you need help choosing the right device?Our sales engineers are ready to guide you to the device that best fits your application. How to contact us?You can use our quote function or contact form with no obligation. With the quote function, you simply add the devices you are interested in and receive a quote. We will get back to you within 24 hours on business days.
After you orderWhen your order is placed, you will receive a confirmed delivery date within two business days. As soon as your package has been shipped, you will receive an email with tracking details, so you always know where your delivery is. |
![]() |
State-of-the-art manufacturing Our 8,500 sqm integrated and automated manufacturing campus in Denmark covers the entire value-chain from design and development to manufacturing. It allows us to design and optimize for testing and manufacturing thereby constantly driving quality up and costs down. |
![]() |
Dedicated Presale / Aftersales
|
![]() |
Product reliability |
![]() |
Product deployment
|
![]() |
Smart products |